System |
Automatic high pressure steam sterilization |
Operating temperature |
Sterilize |
105 to 135°C |
Liquefy |
60 to 110°C |
Retain |
45 to 60°C |
Preheat |
45 to 80°C |
Dry |
135 to 150°C |
Operating Ambient Temp. |
5 to 35°C |
Maximum pressure capacity |
0.255MPa |
Heating |
Sterilize Pipe |
1000W ×2 |
Drying Pipe |
110V/295W×2, 110V/455W×2 |
110V/295W×2, 110V/455W×2 |
110V/275W ×2, 110V/625W ×2 |
110V/275W ×2, 110V/625W ×2 |
Temp. controller |
PlD controlled by microprocessor |
Temp. setting / display |
Touch panel |
Timer / Timer resolution |
Range: 0 or 1min to 99h59min / 1 min. |
Safety Device |
Sterilize sensor error, sterilize SSR short circuit, dry sensor error, dry SSR short circuit, sterilize heater disconnection, dry heater disconnection, water level detection (liquid expansion method), independent chamber overheat protection, cover unlock error, chamber over pressure protection, under pressure protection, warning about setting error in cooling water container, memory error, pressure switch (0.25MPa), pressure safety valve (0.255MPa) |
Internal dimensions |
⌀370 x D470 mm |
⌀370 x D470 mm |
⌀370 x D750 mm |
⌀370 x D750 mm |
External dimensions WxDxH |
520 x 660 x 881 mm |
520 x 660 x 881 mm |
520 x 660 x 116 mm |
520 x 660 x 1161 mm |
Internal capacity |
50L |
50L |
80L |
80L |
Power source (50/60Hz) |
Voltage |
AC100-120V (no plug, round terminal) |
AC200-240V (no plug, round terminal) |
AC100-120V (no plug, round terminal) |
AC200-240V (no plug, round terminal) |
Sterilize current |
25-21A |
12.5-10.5A |
25-21A |
12.5-10.5A |
Dry current |
13.5A |
8.0A |
15.0A |
9.0A |
Weight |
~113kg |
~113kg |
~137kg |
~137kg |
Included items |
2 pcs. stainless steel mesh basket (⌀332 x D195.5 mm) OSQ-90 |
3 pcs. stainless steel mesh basket (⌀332 x D195.5 mm) OSQ-90 |
Drain board x 1, drain bottle x 1, chemical indicator 1 set, filter x 1, droplet tray x 1 |